one) Fill in the use document in time immediately after operating the glove box, such as "user, water oxygen value, what to do"


2) When replacing gasoline cylinders in the glove box, the pressure is set to the default price ± 3mbar


3) The everyday working pressure is set according to the functioning comfort and ease. When not in use for a long time, the stress is established to one-6mbar


4) Routinely look at the wear of the gloves, check out the oil amount of the vacuum pump (amongst the upstream and downstream markings), and include the pump oil appropriately


5) Clean the drinking water probe at the time just about every 2 to 3 months to be certain accurate detection. (Also need to refer to the precise condition of the environment in diverse person containers


6) If you typically use natural and organic solvents in the vac glove box box, it is advisable that you often change the natural and organic solvent adsorbent material, not later on than 1 yr at the most current


7) Switch the filter element frequently, no later than two decades


eight) In accordance to the true use of your glove box, make your mind up to switch the vacuum pump oil cycle, this sort of as black and yellow pump oil, h2o or much more natural solvents in the pump oil. The oil alternative cycle should not exceed 2 many years at the hottest


nine) If a massive amount of money of organic solvent is inevitably uncovered all through the experiment, the cycle desires to be stopped for the experiment. Right after the experiment, to start with thoroughly clean and then begin the cycle


10) When the ecosystem in the box is poisonous or harmful fuel, the exhaust port of the vacuum pump, the exhaust port of the regeneration gasoline and the exhaust port of the cleaning valve should really be exhausted to the outdoor with an exhaust pipe (when exhaust is not permitted to be outdoor, the exhaust gas treatment method should be carried out first).